User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Empro HA-100-50:100A Shunt DATE: 02-May-96 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 3 NUMBER OF LINES: 80 CONFIGURATION: Fluke 5700A CONFIGURATION: Fluke 5725A CONFIGURATION: Datron 1281 STANDARD: Leeds & Northrup 4223-B STANDARD: Thermo Electric MM II (1) STANDARD: Thermo Electric MM II (2) ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P M F W 1.002 ASK+ X 1.003 HEAD ÍÍ INITIAL CONDITIONS ÍÍ 1.004 DISP The following standard test equipment is needed: 1.004 DISP Thermo Electric Micromite II 1.004 DISP Leeds & Northrup 4223-B 1.005 STD Leeds & Northrup 4223-B 1.006 STD Leeds & Northrup 4223-B 1.007 OPBR Is the MM II's DTB number 25-30? 1.008 JMPF 1.011 1.009 STD Thermo Electric MM II (1) 1.010 JMP 1.012 1.011 STD Thermo Electric MM II (2) 1.012 HEAD ÍÍ INITIAL CONDITIONS ÍÍ 1.013 DISP Connect the HI CURRENT OUTPUT from the Fluke 5725A 1.013 DISP to one of the large current terminals on the .001ê 1.013 DISP Leeds & Northrup Standard Resistor 1.014 DISP Connect the other current terminal on the standard 1.014 DISP resistor to one of the current terminals on the UUT. 1.015 DISP Connect the other current terminal on the UUT 1.015 DISP to the LO CURRENT OUTPUT on the 5725A. 1.016 DISP Connect the potential terminals on the Leeds & Northrup 1.016 DISP resistor to the HI and LO front inputs on the 1281 DMM. 1.017 DISP Connect the potential terminals on the UUT to the 1.017 DISP rear CH B input on the 1281 DMM. 1.018 DISP Insert temperature probe into the center well 1.018 DISP of the standard resistor. 1.019 IEEE [@1281]*RST 1.020 IEEE [@1281]DCV AUTO;RESL7 1.021 5700 3A B1 S 2W 1.022 DISP Observe the Datron 1281, until the millivolt 1.022 DISP reading stabilizes. 1.023 CALL Standard Resistor Correction 1.024 MEME >3 1.025 HEAD {} 1.026 HEAD {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 1.027 HEAD {º Shunt Verification º} 1.028 HEAD {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 1.029 HEAD ÍÍ Shunt Verification ÍÍ 1.030 JMP 2.001 1.031 EVAL 2.001 IEEE [@1281]*RST 2.002 IEEE [@1281]DCV AUTO 2.003 5700 3A B1 S 2W 2.004 IEEE [D3000] 2.005 IEEE [@1281]RDG?[I] 2.006 MEME >1 2.007 IEEE [@1281]INPUT CH_B 2.008 IEEE [@1281]RDG?[I] 2.009 MEME >2 2.010 5700 * S 2.011 MATH m[1]=m[1]/m[3] 2.012 MATH m[2]=m[2]/m[1] 2.013 MATH mem=.5 2.014 MATH mem1= m[2] 2.015 MATH mem1=mem1*1000 2.016 MEMC mZ 0.25000% ÿ3A 3.001 IEEE [@1281]*RST 3.002 HEAD {} 3.003 DISP {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 3.003 DISP {º THIS COMPLETES THE VERIFICATION º} 3.003 DISP {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 3.004 END